Technická specifikace
Specifikace:
Full Scale Range:1, 2, 10, 100, 1000, 5000, 10000, 15000, 20000, and 25000 Torr (mmHg)
Resolution:
0.01% of F.S.
Accuracy:
0.5% of Reading
(± temperature coefficients)
Temperature Coefficients:
Zero 0.02% F .S./°C
Span 0.04%ofRdg.
Ambient Operating TemperatureSensor
0° to 150°C (bakeable to 200°C with cable removed)
PreAmp
0° to 50°C
Overpressure Limit120% of Sensor F.S. or 35 psia whichever is greater
Materials Exposed to Gases
Inconel and 316 SST
Volume
7.0 cc
Input Power Required:
±15 VDC @ 35mA regulated to ±5%
Output0 to 10 VDC into >10kΩ
Electromagnetic Compatibility:
Fully compliant to EMC Directive 2004/108/EC when used with an overall metal braided
shielded cab
le properly grounded at both ends
Electrical Connector:
14-pin Micro
Fittings:
0.50 inch (12.7 mm)
O.D.tube(standard),
8 femaleVCR®
1.33 inch (33.8 mm) Conflat®
NW16-KF