| Full Scale Range | 100 Torr |
| Resolution | 0.01% of Full Scale |
| Accuracy | 0.5% of Reading (± temperature coefficients) |
| Temperature Coefficients | |
| Zero | 0.02% Full Scale/°C |
| Span | 0.04% of Reading. |
| Ambient Operating Temperature | |
| Sensor | 0° to 150°C (bakeable to 200°C with cable removed) |
| PreAmp | 0° to 50°C |
| Overpressure Limit | 120% of Sensor Full Scale or 35 psia whichever is greater |
| Materials Exposed to Gases | Inconel and 316 SST |
| Volume | 7.0 cc |
| Input Power Required | ±15 VDC @ 35mA regulated to ±5% |
| Output | 0 to 10 VDC into >10k Ω |
| Compliance | Fully CE Compliant to EMC Directive 2004/108/EC (when used with an overall metal braided shielded cable properly grounded at both ends) |
| Electrical Connector | 14-pin Micro |
| Fittings | 8 VCR® female |