Full Scale Range | 100 Torr |
Resolution | 0.01% of Full Scale |
Accuracy | 0.5% of Reading (± temperature coefficients) |
Temperature Coefficients | |
Zero | 0.02% Full Scale/°C |
Span | 0.04% of Reading. |
Ambient Operating Temperature | |
Sensor | 0° to 150°C (bakeable to 200°C with cable removed) |
PreAmp | 0° to 50°C |
Overpressure Limit | 120% of Sensor Full Scale or 35 psia whichever is greater |
Materials Exposed to Gases | Inconel and 316 SST |
Volume | 7.0 cc |
Input Power Required | ±15 VDC @ 35mA regulated to ±5% |
Output | 0 to 10 VDC into >10k Ω |
Compliance | Fully CE Compliant to EMC Directive 2004/108/EC (when used with an overall metal braided shielded cable properly grounded at both ends) |
Electrical Connector | 14-pin Micro |
Fittings | 8 VCR® female |