Full Scale Ranges | 10, 100, 1000, 5000, 10000, 15000, 20000, 25000 mmHg |
Resolution | 0.01% of Full Scale |
Accuracy | 0.5% of Reading (± temperature coefficients) |
Temperature Coefficients | |
Zero | 0.02% of Full Scale/°C |
Span | 0.04% of Reading/°C |
Ambient Operating Temperature | |
Sensor | 0° to 150°C (bakeable to 200°C with cable removed) |
Preamp | 0° to 50°C |
Maximum Overpressure | 120% of Full Scale or 20 psi (140 kPa), whichever is greater (for Full Scale ranges less than 1000 mmHg, the following restrictions apply: if high pressure is on the Measurement Side (Px) of the sensor, the maximum differential overpressure allowable is 120% of Full Scale or 20 psi, whichever is greater; if high pressure is on the Reference Side (Pr) of the sensor, the maximum differential overpressure allowable is 120% of Full Scale Consult factory for higher Pr overpressure protection.) |
Maximum Line Pressure | 40 psig (275 kPa) |
Materials Exposed to Gases | |
Measurement Side (Px) | Inconel |
Reference Side (Pr) | Inconel, Ceramic, Palladium, Stainless Steel, Glass |
Volume | |
Measurement Side (Px) | 7.9 cc |
Reference Side (Pr) | 19.0 cc |
Fittings | |
Standard | ½ inch (12.7 mm) tubulation |
Optional | Swagelok® 8 VCR® female, NW 16 KF, mini-CF rotatable (high pressure units, 5000 to 25000 mmHg Full Scale, supplied with 8 VCR fittings only) |
Power Required | ±15 VDC @ 35 mA regulated ±2% |
Output | 0 to 10 VDC into > 10K Ω load |
Electrical Connector | AMP 14-pin ribbon connectors; MKS Connector Kit, 221B-K1 |
Compliance | Fully compliant to CE Directive 2004/108/EEC |